# Cycle data: sampling, storage, audit trail

> A 100 Hz record is not free. How to choose a rate per test, keep the raw data intact, and hand a defensible record to the LIMS.

Canonical: https://batteryxlab.com/notes/cycle-data-management/
Language: en

- [Home](https://batteryxlab.com/)
- Resources
- [Notes](https://batteryxlab.com/notes/)
- Cycle data: sampling, storage, audit trail

Measurement · Service

# Cycle data: sampling, storage, audit trail

A 100 Hz record is not free. How to choose a rate per test, keep the raw data intact, and hand a defensible record to the LIMS.

19 Nov 2026 · 8 min read · [All notes](https://batteryxlab.com/notes/)

![BTS-6008A-5V300A-H]()
BTS-6008A-5V300A-H

Cycle data is easy to create and hard to defend. A tester can write millions of samples over a long campaign, and the file size is often the first thing a team notices. The more important questions come later: which samples were kept, whether the raw record can still be reconstructed, whether the audit trail explains a method change, and whether the LIMS record points back to the channel data or only to a summary. This note covers the choices that decide those answers.

## The data path is a chain, not a file

A cycle record passes through several systems before it is reviewed. The tester samples voltage, current, time and temperature on each channel. The control software applies the step file, records events and stores the result. A database holds the records; an export or a LIMS receives the summary; a report is written from the summary. Each hand-off is a place where data can be lost, re-keyed or detached from its context.

The site documentation describes the software as a single application for instrument control, LIMS workflow and analysis, with an append-only audit trail and role-based access. The formation page describes per-cell records linked to module and pack serials, with LIMS-friendly export. Those are the intended properties. They still depend on the test plan recording the right fields and on the export preserving the link to the raw data.

Datasheet fact: the BTS-6008A-5V300A-H publishes a recording frequency of 100 Hz, a minimum time interval of 10 ms, a minimum voltage interval of 10 mV and minimum current intervals of 150 mA, 300 mA and 600 mA on its three ranges. The BTS-4008Q-5V100mA-124 publishes 10 Hz, a 100 ms minimum record interval, a 5 mV minimum voltage change and current change thresholds from 0.1 µA to 100 µA depending on range. Both publish MySQL storage and power-off data protection. The BTS-6008A adds CSV/Excel/TXT export, barcode scanning and historical-data traceability; the BTS-4008Q-5V100mA publishes Excel/TXT export. The recording conditions are not the same as the accuracy specification: they define when a sample is written, not how correct it is.

## Sampling rate is a test decision

Record at the rate the result needs, not at the highest rate the instrument offers. A slow cycle-life test can run at 10 Hz and still capture the capacity trend. A pulse test that needs the ohmic drop at 10 ms needs a channel and a recording interval that can see it. A dQ/dV analysis benefits from a dense voltage record around the plateaus; a long rest step does not.

The cost of a high rate is not only disk space. It is also the time to transfer the record, the time to query it, the size of the backup, and the chance that a review tool samples it down without saying so. A record that is too coarse hides the feature; a record that is too dense hides the feature under its own volume.

Illustrative arithmetic: assume eight channels at 100 Hz, four stored values per sample and eight bytes per value. That is 8 × 100 × 4 × 8 = 25,600 bytes per second, or about 92 MB per hour before database overhead. At 10 Hz the same assumptions give about 9.2 MB per hour. A 1,000-hour campaign is roughly 92 GB at 100 Hz versus 9.2 GB at 10 Hz. The numbers are a planning estimate, not a measurement of the product's file format, but they show why the rate belongs in the test method.

There is a third option between “record everything” and “record nothing”: record at the native rate during the step that needs it, then archive a decimated series for long-term review while keeping the raw segment for the defined retention window. If you decimate, record the algorithm and the factor in the metadata. A decimated file that does not say it was decimated is a trap for the next reviewer.

## What “recording condition” means

The datasheets list minimum time, voltage and current intervals. These are thresholds that decide when a new record is written. A large voltage change can trigger a sample before the minimum time has elapsed; a stable rest step can produce fewer samples than the nominal rate suggests. The consequence is that two tests with the same nominal frequency can have different sample counts and different time gaps.

For integration and dQ/dV, use the recorded timestamps, not an assumed fixed interval. If the analysis assumes one sample per 10 ms and the record contains gaps, the capacity integral is wrong even when every individual sample is correct. Store the actual interval or the timestamp with each sample, and make the analysis tolerate gaps explicitly.

## Data integrity: the audit trail is part of the data

An audit trail is not a log file bolted onto the side of the result. It is the evidence that the result is attributable, contemporaneous and unaltered. The software page lists an append-only change history, role-based access, operator and timestamp on record changes, and electronic records retained with the test. Those controls answer the reviewer's questions: who started the test, who changed the method, who paused it, and what the record looked like before the change.

Engineering judgement: treat method versions as part of the data. A cycle result without the step file version, the calibration status of the channel and the software version is difficult to reproduce. Record the method name and version, the channel and rack, the calibration certificate or due date, the operator, the start and end times, and any manual intervention. If a step was skipped or repeated, the event should be in the record, not only in an operator's notebook.

Clock discipline matters more than it looks. If the tester, the chamber and the LIMS use different clocks, an event that happened at one moment appears at three. Use a common time source where the architecture allows it, record the offset, and do not “fix” timestamps during export. A corrected timestamp without a record of the correction is worse than an offset that is documented.

## LIMS export: link, do not duplicate

The LIMS owns sample identity, test assignment, result status and retention policy. The tester owns the raw channel record. The export should connect the two, not replace either. A useful export carries the sample ID, the test ID, the channel and rack, the method version, the calibration reference, the summary result with units, the status, and a pointer or hash to the raw record.

Two-way integration is described on the software page: sample ID, test assignment, result and status. The practical rule is that the LIMS should not become the raw waveform store, and the tester should not become the master of sample identity. If the same cell ID is typed into both systems, one of them will eventually disagree. Use the barcode or the LIMS identifier as the key, and reject a test start that has no key.

Summary tables are useful for review and useless for re-analysis if the raw data is gone. Keep the step boundaries, the event log and the channel-level summary alongside the raw record. A reviewer should be able to move from a capacity fade point in the summary to the cycle that produced it, and from that cycle to the samples and the method version that generated it.

## Failure modes to plan for

- **Disk full or database unavailable.** Define what the tester does when it cannot write. Power-off data protection helps with a power loss, not with a full volume.

- **Network loss.** If the LIMS is unreachable, the test should continue locally and queue the export; it should not silently drop the result.

- **Duplicate or misread identifiers.** A barcode misread can attach a result to the wrong cell. Validate the identifier against the LIMS before current is applied.

- **Clock drift.** Record the offset and align events in the analysis, not by editing the source timestamps.

- **Export without metadata.** A CSV with voltage and current but no channel, method or units is not a record; it is a plot waiting to be misread.

## Takeaways

- Choose the recording rate per test: 100 Hz on the BTS-6008A and 10 Hz on the BTS-4008Q-5V100mA are different tools for different features, and the minimum intervals decide what the record can resolve.

- Keep the audit trail, method version, calibration reference and channel identity with the data; an exported summary without them is not reproducible.

- Let the LIMS own sample identity and the tester own raw data, and connect them with a stable key, a pointer and a hash rather than re-keying the result.

## Request a configuration review

Tell us the channel count, current range and protocol. We will check the measurement class against the note and return a configuration, not a generic datasheet.

[Request a configuration review](https://batteryxlab.com/contact/?note=cycle-data-management)

## References

- [BTS-6008A-5V300A-H](https://batteryxlab.com/products/bts-6008a-5v300a-h/) — [datasheet (PDF)](https://batteryxlab.com/assets/docs/BTS-6008A-5V300A-H.pdf)

- [BTS-4008Q-5V100mA-124](https://batteryxlab.com/products/bts-4008q-5v100ma-124/) — [datasheet (PDF)](https://batteryxlab.com/assets/docs/BTS-4008Q-5V100mA-124.pdf)

- [BTS-4008Q-5V20A-SR](https://batteryxlab.com/products/bts-4008q-5v20a-sr/) — [datasheet (PDF)](https://batteryxlab.com/assets/docs/BTS-4008Q-5V20A-SR.pdf)

- [BTS-4008 / 6000A / CA series](https://batteryxlab.com/series/bts-4008/)

- [BTS-4008Q series](https://batteryxlab.com/series/bts-4008q/)

- [EV battery testing](https://batteryxlab.com/industries/ev-battery-testing/)

- [BESS testing](https://batteryxlab.com/industries/bess-testing/)

- Peer-reviewed use of this instrument class: [Autonomous optimization of non-aqueous Li-ion battery electrolytes via robotic experimentation and machine learning coupling](https://doi.org/10.1038/s41467-022-32938-1) — Dave (2022), Nature Communications. DOI: 10.1038/s41467-022-32938-1

- [BTS 9.1 test software](https://batteryxlab.com/software/)

- [Formation & grading records](https://batteryxlab.com/formation-grading/)

- [Datasheets and glossary](https://batteryxlab.com/resources/)

[← All technical notes](https://batteryxlab.com/notes/)