# DCIR pulse timing: where the numbers come from

Current response, sampling interval and range selection decide whether a DCIR value is reproducible. A timing budget for 10 ms and 10 s measurements.

8 Nov 2026 · 5 min read · [All notes](https://batteryxlab.com/notes/)

DCIR is often reported as a single number with no timing attached. That number is not reproducible. The same cell can give different DCIR values on the same instrument if the voltage is sampled 10 ms after the current step instead of 10 s, if the current actually delivered differs from the setpoint, or if the rest period before the pulse was too short. This note sets out the timing chain from command to recorded sample and shows where the common errors enter.

## Define the two points before choosing hardware

DCIR is ΔV/ΔI between two defined states. The first state is the settled open-circuit voltage before the pulse. The second is the loaded voltage at a stated time after the current step — 10 ms for an ohmic check, 10 s for a conventional DCIR, or the end of the pulse for a protocol-defined value. Write both times into the test method. “DCIR” without a time is not a complete specification.

The current step also needs a definition. Command time and current-reached time are not the same. The BTS-6008A-5V300A-H publishes a current response of ≤3 ms; the BTS-4008Q-5V100mA-124 publishes ≤1 ms (10% to 90%). Start the ΔV window after the measured current has settled, not at the software command. The recorded current channel is the evidence: if it has not reached the setpoint, the voltage is still moving for reasons that have nothing to do with the cell.

## The sampling chain sets the earliest usable point

On the BTS-6008A-5V300A-H, recording is 100 Hz — one sample every 10 ms — with a minimum recording interval of 10 ms. The minimum pulse width is 50 ms and the minimum step time is 0.1 s. On the BTS-4008Q-5V100mA-124, recording is 10 Hz, the minimum recording interval is 100 ms and the minimum pulse width is 500 ms. These are different instruments for different pulse lengths, and the recording rate decides which part of the transient you can actually see.

| Interval to measure | BTS-6008A-5V300A-H | BTS-4008Q-5V100mA-124 | Practical consequence |
| --- | --- | --- | --- |
| Current response | ≤3 ms | ≤1 ms | Do not take a sample before the current has settled |
| Recording interval | 10 ms (100 Hz) | 100 ms (10 Hz) | Ohmic drop at 10 ms is visible only on the 100 Hz channel |
| Minimum pulse | 50 ms | 500 ms | A 10 ms ohmic pulse is below the published minimum on both |
| Typical 10 s DCIR | ~1,000 samples | ~100 samples | Both can resolve the endpoint; only one resolves the first 100 ms |

Timing values are from the published datasheets. Sample counts assume a continuous 10 s pulse and no gaps.

## Use the measured current, not the nominal current

ΔI is the current that actually flowed. Current accuracy is published as a percentage of full scale, so the error depends on which range the channel selected. On the BTS-6008A-5V300A-H, ±0.05% of the 300 A range is ±150 mA. A 10 A pulse measured on that range carries a 1.5% current uncertainty before any voltage error is considered. The 75 A range reduces the same specification to ±37.5 mA, or 0.375% of a 10 A pulse. Range selection is therefore part of the DCIR method, not an implementation detail.

The voltage side has the same structure. The BTS-6008A publishes ±0.02% of full scale on a 5 V range, which is ±1 mV. If the loaded voltage moves by 20 mV during the pulse, a 1 mV error is 5% of the result. Averaging several samples around the endpoint helps with random noise, but it does not remove a gain error or a range offset. Calibration and a known reference cell are the checks for those.

## Pulse length, rest and temperature

A DCIR pulse heats the cell and concentrates lithium at the electrode surface. The rest period before the pulse has to return the cell to a defined state; the rest after the pulse has to be long enough that the next pulse does not start from a different polarisation. Ten seconds of pulse and ten seconds of rest is a common starting point, but the correct values belong to the cell specification and the comparison you want to make. Keep them fixed across the whole campaign.

Temperature is not a minor correction. Electrolyte conductivity, interfacial kinetics and contact resistance all move with temperature, and a cell that is 2 °C warmer than its neighbour will show a lower DCIR for reasons that are not electrochemical. Use the chamber setpoint, allow a soak that is long enough for the cell mass, and record the chamber temperature and the cell-surface temperature if a sensor is fitted. Do not compare a 25 °C pulse with a 23 °C pulse and call the difference degradation.

## A repeatable sequence

- Rest at the target temperature until the open-circuit voltage drift is below the resolution you care about.

- Select the current range closest to the pulse current, then confirm the range in the data header.

- Record the pre-pulse voltage over a stable window; use the mean, not one sample.

- Apply the pulse, find the point where the measured current first reaches the setpoint, and start the timing window there.

- Take the loaded voltage at the defined time, using the mean of the samples inside a short window around it.

- Compute ΔV/ΔI with the measured current; report the time, the range and the temperature with the result.

Run the sequence three times on a reference cell; if the results disagree beyond the uncertainty you need, the method or fixture is the problem.

## Takeaways

- DCIR is only defined with a time: state whether the voltage is taken at 10 ms, at the end of a 10 s pulse, or at another protocol-defined point.

- Start the measurement window when the measured current has settled, not when the command is sent; the published response is ≤3 ms on the BTS-6008A and ≤1 ms on the BTS-4008Q.

- Choose the current range for the pulse, not for the test programme: a 10 A pulse on the 300 A range carries ±150 mA of current uncertainty.

## Request a configuration review

Tell us the channel count, current range and protocol. We will check the measurement class against the note and return a configuration, not a generic datasheet.

[Request a configuration review](https://batteryxlab.com/contact/?note=dcir-pulse-timing)

## References

- [BTS-6008A-5V300A-H](https://batteryxlab.com/products/bts-6008a-5v300a-h/) — [datasheet (PDF)](https://batteryxlab.com/assets/docs/BTS-6008A-5V300A-H.pdf)

- [BTS-4008Q-5V100mA-124](https://batteryxlab.com/products/bts-4008q-5v100ma-124/) — [datasheet (PDF)](https://batteryxlab.com/assets/docs/BTS-4008Q-5V100mA-124.pdf)

- [BTS-4008 / 6000A / CA series](https://batteryxlab.com/series/bts-4008/)

- [BTS-4008Q series](https://batteryxlab.com/series/bts-4008q/)

- [EV battery testing](https://batteryxlab.com/industries/ev-battery-testing/)

- [BESS testing](https://batteryxlab.com/industries/bess-testing/)

- Peer-reviewed use of this instrument class: [Expediting Lithium Electrochemistry via a Bilayer for High-Rate Lithium Metal Batteries](https://doi.org/10.1007/s40820-026-02146-3) — Park (2026), Nano-Micro Letters. DOI: 10.1007/s40820-026-02146-3

- [Applications: fast-charge and EV profiles](https://batteryxlab.com/applications/)

[← All technical notes](https://batteryxlab.com/notes/)

## Contact

- Email: sales@batteryxlab.com
- Phone: +852 4436 9298
- WhatsApp: https://wa.me/85244369298
- Request a quote: https://batteryxlab.com/contact/
- Full site index for agents: https://batteryxlab.com/llms.txt