Close-up of cylindrical cells connected to a test channel module

Materials research

Coin cells, microamp currents and reproducible screening

Materials screening compares nominally identical cells. The instrument has to resolve the difference between two materials without adding a difference of its own.

Test challenges

What makes this application difficult

The signal is small

Coin-cell leakage and low-rate cycling sit in the microamp range. A channel whose lowest range starts at 0.5 µA covers that work without adding a separate picoammeter and a second data stream.

Contact resistance is part of the result

Coin-cell holders add resistance and a temperature gradient. Four-terminal sensing at the holder and a consistent assembly torque keep cell-to-cell comparisons meaningful.

Temperature changes the ranking

Arrhenius behaviour means a 2 °C offset can reorder two candidate materials. Chamber stability and a recorded temperature channel belong to the method, not to the facilities checklist.

Three cells are not statistics

Screening needs enough channels per variant to separate a real effect from assembly variance. Eight channels per unit, twenty units per rack, keeps a materials matrix inside one schedule.

Recommended configuration

A configuration that works as a chain

Every model below is a published datasheet in the model library. Channel count, ranges and accuracy are read from that datasheet at build time, not retyped here.

Cell · BTS-4008Q

BTS-4008Q-5V100mA-124

Four current ranges from 0.5 µA to 100 mA, with 0.01% voltage accuracy. Eight channels per unit and up to twenty units per rack for parallel coin-cell screening.

Channels
8
Voltage
10mV~5V
Current ranges
R1: 0.5μA~0.1mA · R2: 0.1mA~1mA · R3: 1mA~10mA · R4: 10mA~100mA
Accuracy
V ±0.01% of FS · I ±0.02% of FS
Sampling
10Hz
Dimensions
483mm x 310mm x 48mm

Open datasheet →

BTS-4008Q-5V6A-S1/164

0.5 mA to 6 A over four ranges for pouch and small cylindrical cells, using the same schedule format as the microamp unit.

Channels
8
Voltage
25mV~5V
Current ranges
R1: 0.5mA~100mA · R2: 100mA~500mA · R3: 500mA~3A · R4: 3A~6A
Accuracy
V ±0.02% of FS · I ±0.02% of FS
Sampling
10Hz
Dimensions
483mm x 404mm x 130mm

Open datasheet →

BT-9562

AC four-terminal internal resistance at 1 kHz, four ranges from 3.3 mΩ to 3.3 Ω, to separate electrode and electrolyte contributions from contact resistance.

Impedance ranges
-3.3~3.3mΩ · -33~33mΩ · -330~330mΩ · -3.3~3.3Ω
Impedance accuracy
±0.4% · ±0.3% · ±0.3% · ±0.3%
Test method
AC 4-terminal method (1kHz ±0.02Hz)
Dimensions
W220mm × H80mm × D300mm
Weight
2.5kg
Input power
45VA

Open datasheet →

Typical sequence

  1. Electrode and slurry prep

    Coating weight, calendering and punch diameter recorded per batch; the cell build is part of the experiment, not a prelude to it.

  2. Coin-cell assembly

    Consistent separator, electrolyte volume and crimping torque; holders are mapped to channels before the first cycle.

  3. Formation

    Low-rate first cycles with voltage limits and a recorded coulombic efficiency for every cell.

  4. Rate capability and impedance

    Stepped C-rates at controlled temperature, then AC internal resistance on the same cells without remounting.

  5. Cycling and report

    Long-term cycling at the chosen rate, with capacity, efficiency and DCIR exported per cell and per material variant.

Records that survive an audit

A materials result has to survive review by someone who was not in the lab. Method, instrument and calibration need to be in the same record.

  • BTS 9.1 stores the full schedule and measured values, with an audit trail for edits.
  • CNAS calibration to ISO/IEC 17025 is issued per module and referenced in the method record.
  • LIMS import and export keep cell ID, material variant and build batch together.
  • Exports include the sampling rate and current range in use, so rate-dependent results can be interpreted.

Delivery lead time

Working days from a confirmed order. The same table and conditions as the solutions page.

Delivery lead time by product type and destination
TypeMainland ChinaOverseas
Standard products 40 working days 80 working days
Non-standard products Standard lead time + 50% (i.e. 60 / 120 working days) Same
  • Counted in working days; weekends and statutory holidays are not included.
  • Counted from order confirmation and receipt of the advance payment.
  • Sea or air freight and customs clearance are not included.
  • The final lead time is confirmed in the quotation / contract.

Set up a materials screen

Send the cell format, current range, channel count and temperature requirements. We reply with a configuration and the fixtures needed to make it repeatable.

Talk to an engineer